Instrument Maintenance

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

191

6.3. PREDICTIVE DIAGNOSTICS

The analyzer’s test functions can be used to predict failures by looking at trends in their
values. Initially it may be useful to compare the state of these test functions to the
values measured on your instrument at the factory and recorded on the 9110TH/M Final
Test and Validation Data Form
(TAI part number 04490, attached to the manual). Table
6-2 can be used as a basis for taking action as these values change with time. The
internal data acquisition system (DAS) is a convenient way to record and track these
changes. APICOM control software can be used to download and review these data
even from remote locations (Section 4.15.2.8 describes APICOM).

Table 6-2: Predictive Uses for Test Functions

FUNCTION EXPECTED

ACTUAL

INTERPRETATION

&

ACTION

RCEL

pressure

Constant to
within ± 0.5

Fluctuating

Developing leak in pneumatic system. Check for leaks

Slowly increasing

Pump performance is degrading. Replace pump head
when pressure is above 10 in-Hg-A

SAMPLE

pressure

Constant within

atmospheric

changes

Fluctuating

Developing leak in pneumatic system. Check for leaks

Slowly decreasing

Flow path is clogging up. Replace orifice filters

Slowly increasing

Developing leak in pneumatic system to vacuum
(developing valve failure). Check for leaks

Ozone Flow

Constant to

within ± 15

Slowly decreasing

Flow path is clogging up. Replace orifice filters

AZERO

Constant within

±20 of check-out

value

Significantly

increasing

Developing AZERO valve failure. Replace valve
PMT cooler failure. Check cooler, circuit, and power
supplies
Developing light leak. Leak check.
O

3

air filter cartridge is exhausted. Change chemical

NO2 CONC

Constant for

constant

concentrations

Slowly decreasing

signal for same

concentration

Converter efficiency may be degrading. Replace
converter.

NO CONC

Constant for

constant

concentration

Decreasing over time

Drift of instrument response; clean RCEL window,
change O

3

air filter chemical.

6.4. MAINTENANCE PROCEDURES

The following procedures need to be performed regularly as part of the standard
maintenance of the Model 9110TH/M.

6.4.1. CHANGING THE SAMPLE PARTICULATE FILTER

The particulate filter should be inspected often for signs of plugging or excess dirt. It
should be replaced according to the service interval in Table 9-1 even without obvious
signs of dirt. Filters with 1 µm pore size can clog up while retaining a clean look. We
recommend to handle the filter and the wetted surfaces of the filter housing with gloves