Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 155

Advertising
background image

Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

135

To manually adjust the signal levels of an analog output channel, press:

From the

AIO CONFIGURATION SUBMENU

(See Section 6.13.4.1)

DIAG

ANALOG I/O CONFIGURATION

PREV NEXT

ENTR

EXIT

Continue pressing SET> until you reach the

output to be configured

DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL

<SET SET>

EDIT

EXIT

Continue pressing SET> until ...

DIAG AIO DATA_OUT_2 CALIBRATED:NO

<SET SET>

CAL

EXIT

DIAG AIO DATA_OUT_2 RANGE: 5V

SET>

EDIT

EXIT

DIAG AIO DATA_OUT_2 VOLT-Z: 0 mV

U100 UP10 UP DOWN DN10 D100 ENTR

EXIT

DIAG AIO DATA_OUT_2 VOLT-S: 4500 mV

U100 UP10 UP DOWN DN10 D100 ENTR EXIT

DIAG AIO DATA_OUT_2 CALIBRATED: YES

<SET SET> CAL

EXIT

These menus
only appear if

AUTO-CAL is

turned OFF

These buttons increase /

decrease the analog output

signal level (not the value on the

display)

by 100, 10 or 1 counts.

Continue adjustments until the

voltage measured at the output
of the analyzer and/or the input

of the recording device matches

the value in the upper right hand

corner of the display (within the

tolerances

listed in Table 6-24).

DIAG AIO AOUTS CALIBRATED: NO

SET>

CAL

EXIT

4.13.6.3. Manual Calibration of Analog Outputs Configured for Current Loop Ranges

The current loop output option (see Section 5.4) uses a small converter assembly to
change the DC voltage output by the standard voltage output to a current signal ranging
between 0-20 mA. Since the exact current increment per voltage count varies from
converter to converter and from instrument to instrument, analog outputs with this
option installed cannot be calibrated automatically and must be adjusted manually.

Adjusting the signal zero and full scale values of the current loop output is done in a
similar manner as manually adjusting analog outputs configured for voltage output
except that:

 In this case calibration is performed with a current meter connected in series

with the output circuitry (See Figure 4-12).

 Adjustments to the output are made using the front panel touch screen, also

in 100, 10 or 1 count increments, but the change in the voltage driving the
converter assembly is displayed on the front panel.