Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 249

Advertising
background image

Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

229

Functions of the Relay PCA include:

 handling all AC and DC power distribution including power to the pump.

 a set of jumpers that connect AC power to heaters included in several

optional items, such as the zero/span valve options and the O

2

sensor option

available on the 9110TH/M analyzers.

7.5.4.1. AC Configuration – Internal Pump (JP7)

AC power configuration for internal pumps is set using Jumper set JP7 (see Figure 7-4
for the location of JP7).

Table 7-3:

AC Power Configuration for Internal Pumps (JP7)

LINE

POWER

LINE

FREQUENCY

JUMPER

COLOR

FUNCTION

JUMPER

BETWEEN

PINS

110VAC

115 VAC

60 HZ

WHITE

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

Connects pump pins 2 & 4 to Neutral

4 to 9

50 HZ

1

BLACK

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

Connects pump pins 2 & 4 to Neutral

4 to 9

220VAC

240 VAC

60 HZ

BROWN

Connects pump pins 3 and 4 together

1 to 6

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

50 HZ

1

BLUE

Connects pump pins 3 and 4 together

1 to 6

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

1

A jumper between pins 5 and 10 may be present on the jumper plug assembly, but has no function on the 9110TH/M

analyzers.