Turning an analog output on or off – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 150
Model 9110TH NOx Analyzer
Operating
Instructions
Teledyne Analytical Instruments
130
4.13.5.10. Turning an Analog Output On or Off
Each output can be temporarily turned off. When off, no data is sent to the output.
Electronically, it is still active, but there is simply no data being output, so the signal
level at the rear of the instrument will fall to zero.
To turn an individual analog output channel ON/OFF, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 OUTPUT: ON
<SET SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 OUTPUT: 5V
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 OUTPUT: ON
ON
ENTR EXIT
DIAG AIO DATA_OUT_2 OUTPUT: OFF
OFF
ENTR
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Toggle this
button to turn
the channel
ON/OFF
Continue pressing SET> until ...
Continue pressing SET> until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT