Page 182: 213 o
Model 9110TH NOx Analyzer
Operating
Instructions
Teledyne Analytical Instruments
162
4.15.4.5. Hessen Protocol Gas ID
Since the 9110TH/M measures NOx, NO
2
, NO and O
2
(if the optional sensor is
installed), all of these gases are listed in the Hessen protocol gas list. In its default state
the Hessen protocol firmware assigns each of these gases a Hessen ID number and
actively reports all of them even if the instrument is only measuring one (see
MEASURE_MODE, Section 4.12) .
To change or edit these settings press:
BUTTON FUNCTION
<PREV
Moves to nex t gas entry in list
NEXT>
Moves the cursor previous gas entry in list
INS
Inserts a new gas entry into the list.
DEL
Deletes the >>>>>.
ENTR
Accepts the new setting and returns to the prev ious menu.
EXIT
Ignores the new setting and returns to the previous menu.
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXI
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
HESSEN RESPONSE MODE :CMD
<SET
SET>
EDIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
HESN
COM1
COM2
EXIT
SETUP X.X
HESSEN VARIATION: TYPE 1
SET>
EDIT
EXIT
SETUP X.X
HESSEN GAS LIST
<SET
SET>
EDIT
EXIT
ENTR accepts the
new settings
EXIT ignores the
new settings
SETUP X.X
NOX, 211, REPORTED
<PREV
NEXT>
INS
DEL
EDIT
PRNT EXIT
SETUP X.X
GAS TYPE NOX
<PREV
NEXT>
ENTR
EXIT
Use the PREV & NEXT keys to cycle
through available gases
SETUP X.X
GAS ID: 211
0 0 0
ENTR
EXIT
Toggle to change the gas ID number for the
chosen gas.
SETUP X.X
REPORTED : ON
ON
ENTR
EXIT
Toggle to switch reporting Between ON and
OFF
Use the PREV & NEXT keys to cycle
existing entries in Hessen gas list
Table 4-35: 9110TH/M Hessen GAS ID List
GAS DEFAULT
HESSEN GAS ID
NOx 211
NO 212
NO
2
213
O
2
214