Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 124

Advertising
background image

Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

104

Analyzer

Analyzer

Analyzer

Last Analyzer

Female DB9

Male DB9

RS-232

COM2

RS-232

COM2

RS-232

COM2

RS-232

COM2

Host

RS-232 port

Ensure jumper is

installed between

JP2 pins 21

22 in

last instrument of

multidrop chain.

Figure 4-9:

RS-232-Multidrop Host-to-Analyzer Interconnect Diagram

7. BEFORE communicating from the host, power on the instruments and check that

the Machine ID (Section 4.11.1) is unique for each. On the front panel menu, use
SETUP>MORE>COMM>ID. The default ID is typically the model number or “0”; to
change the 4-digit identification number, press the button below the digit to be
changed; once changed, press/select ENTER to accept the new ID for that
instrument.

8. Next, in the SETUP>MORE>COMM>COM1 menu (do not use the COM2 menu for

multidrop), edit the COM1 MODE parameter as follows: press/select EDIT and set
only QUIET MODE, COMPUTER MODE, and MULTIDROP MODE to ON. Do not
change any other settings.

9. Press/select ENTER to accept the changed settings, and ensure that COM1 MODE

now shows 35.

10. Press/select SET> to go to the COM1 BAUD RATE menu and ensure it reads the

same for all instruments (edit as needed so that all instruments are set at the same
baud rate).

NOTES:

The (communication) Host instrument can address only one instrument at a time,
each by its unique ID (see Step 7 above).

TAI recommends setting up the first link, between the Host and the first analyzer,
and testing it before setting up the rest of the chain.

Advertising