Connecting the status outputs – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 36

Advertising
background image

Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

16

Table 3-1: Analog Output Data Type Default Settings

PARAMETER

CHANNEL DEFAULT SETTING

A1 A2 A3

A4

3

DATA TYPE

1

NXCNC1 NOCNC1 N2CNC1

NXCNC2

RANGE

0 - 5 VDC

2

REC OFS

0 mVDC

AUTO CAL.

ON

CALIBRATED NO

OUTPUT ON

SCALE 100

ppm

UPDATE 5

sec

1

See Table A-6 of 9110TH/M Appendix A for definitions of these DAS data types

2

Optional current loop outputs are available for analog output channels A1-A3.

3

On analyzers with O

2

sensor options installed, DAS parameter O2CONC is assigned to output A4.

To access these signals attach a strip chart recorder and/or data-logger to the appropriate
contacts of the analog output connecter on the rear panel of the analyzer.

ANALOG OUT

A1 A2 A3 A4
+ - + - + - + -

Figure 3-7:

Analog Output Connector

Table 3-4: Analog Output Pin-Outs

PIN

ANALOG OUTPUT

VOLTAGE SIGNAL

CURRENT SIGNAL

1

A1

V Out

I Out +

2

Ground

I Out -

3

A2

V Out

I Out +

4

Ground

I Out -

5

A3

V Out

I Out +

6

Ground

I Out -

7

A4

V Out

Not Available

8 Ground

Not

Available

3.4.4. CONNECTING THE STATUS OUTPUTS

The analyzer’s status outputs to interface with a device that accepts logic-level digital
inputs, such as programmable logic controller (PLC) chips, are accessed through a 12
pin connector labeled STATUS on the analyzer’s rear panel.