Model 9110TH NOx Analyzer

Troubleshooting & Repair

Teledyne Analytical Instruments

250

Figure 7-18:

9110TH/M Sensor Assembly

Figure 7-19.

3-Port Reaction Cell Oriented to the Sensor Housing

4. Remove the two connectors on the PMT housing end plate facing towards the front

panel.

5. Remove the end plate itself (4 screws with plastic washers).

6. Remove the dryer packages inside the PMT housing.

7. Along with the plate, slide out the OPTIC TEST LED and the thermistor that

measures the PMT temperature.

8. Unscrew the PMT assembly, which is held to the cold block by two plastic screws.