Calibration method, Set the expected o, Content of ambient air) – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 69

Advertising
background image

Getting Started

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

49

3.7.2.2. O

2

Calibration Method

STEP 1 – Set O

2

span gas concentration :

Set the expected O

2

span gas concentration.

This should be equal to the percent concentration of the O

2

span gas of the selected

reporting range (default factory setting = 20.8%; the approximate O

2

content of ambient

air).

SAMPLE

GAS TO CAL:NOX

NOX

O2

ENTR EXIT

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST >

CAL

SETUP

M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX

<TST TST> ZERO SPAN

CONC

EXIT

SAMPLE

GAS TO CAL:O2

NOX

O2

ENTR

EXIT

M-P CAL O2 SPAN CONC:20.8%

0

2

0

.8

0

ENTR EXIT

The O

2

span concentration value automatically defaults to

20.8 %.

If this is not the the concentration of the span gas being

used, toggle these buttons to set the correct concentration

of the O

2

calibration gases.

EXIT ignores the new

setting and returns to

the previous display.

ENTR accepts the new

setting and returns to

the previous menu.