Changing the no, Converter – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 215

Advertising
background image

Instrument Maintenance

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

195

6.4.4. CHANGING THE NO

2

CONVERTER

The NO

2

converter is located in the center of the instrument, see Figure 3-5 for location,

and Figure 6-3 for the assembly. The converter is designed for replacement of the
cartridge only, the heater with built-in thermocouple can be reused.

1. Turn off the analyzer power, remove the cover and allow the converter to cool.

2. Remove the top lid of the converter as well as the top layers of the insulation until

the converter cartridge can be seen.

CAUTION

THE CONVERTER OPERATES AT 315º C. SEVERE BURNS CAN RESULT IF THE

ASSEMBLY IS NOT ALLOWED TO COOL. DO NOT HANDLE THE ASSEMBLY UNTIL

IT IS AT ROOM TEMPERATURE. THIS MAY TAKE SEVERAL HOURS.

3. Remove the tube fittings from the converter.

4. Disconnect the power and the thermocouple of the converter. Unscrew the

grounding clamp of the power leads with a Phillips-head screw driver.

5. Remove the converter assembly (cartridge and band heater) from the can. Make a

of the orientation of the tubes relative to the heater cartridge.

6. Unscrew the band heater and loosen it, take out the old converter cartridge.

Figure 6-3:

NO

2

Converter Assembly

7. Wrap the band heater around the new replacement cartridge and tighten the screws

using a high-temperature anti-seize agent such as copper paste. Make sure to use
proper alignment of the heater with respect to the converter tubes.

8. Replace the converter assembly, route the cables through the holes in the can and

reconnect them properly. Reconnect the grounding clamp around the heater leads
for safe operation.

Advertising