Analog output calibration – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 151

Advertising
background image

Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

131

4.13.6. ANALOG OUTPUT CALIBRATION

Analog calibration needs to be carried out on first startup of the analyzer (performed in
the factory as part of the configuration process) or whenever recalibration is required.
The analog outputs can be calibrated automatically, either as a group or individually
(Section 4.13.6.1), or adjusted manually (see Section 4.13.6). (Manual calibration should
be used for the 0.1V range or in cases where the outputs must be closely matched to the
characteristics of the recording device. Manual calibration requires the AUTOCAL
feature to be disabled).During automatic calibration the analyzer tells the output
circuitry to generate a zero mV signal and high-scale point signal (usually about 90% of
chosen analog signal scale) then measures actual signal of the output. Any error at zero
or high-scale is corrected with a slope and offset.

To enable or disable the Auto-Cal feature for one output channel, press.

From the

AIO CONFIGURATION SUBMENU

(See Section 6.13.4.1)

DIAG

ANALOG I/O CONFIGURATION

PREV NEXT

ENTR

EXIT

DIAG AIO DATA_OUT_3: 5V, NXCNC1, NOCAL

<SET SET>

EDIT

EXIT

DIAG AIO DATA_OUT_3 AUTO CAL.:ON

<SET SET>

EDIT

EXIT

DIAG AIO DATA_OUT_3 AUTO CAL.:ON

ON

ENTR EXIT

Toggle this button

to turn AUTO CAL

ON or OFF

(OFF = manual

calibration mode).

ENTR accepts

the new setting.

EXIT ignores the

new setting

Continue pressing SET> until ...

DIAG AIO DATA_OUT_3 RANGE: 5V

SET>

EDIT

EXIT

DIAG AIO DATA_OUT_3 AUTO CAL.:OFF

OFF

ENTR EXIT

Continue pressing SET> until you reach the

output to be configured

DIAG AIO AOUTS CALIBRATED: NO

SET>

CAL

EXIT

Note

Channels with current loop output options cannot be calibrated
automatically. Outputs Configured for 0.1V full scale should always be
calibrated manually.