Sensor module, reaction cell – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 301

Advertising
background image

Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

281

In the unlikely event that the flash chip should fail, the analyzer will continue to operate
with just the DOM. However, all configuration information will be lost, requiring the
unit to be recalibrated.

8.5.2. SENSOR MODULE, REACTION CELL

Electronically, the 9110TH/M sensor assembly (see Figure 9-6) consists of several
subassemblies with different tasks: to detect the intensity of the light from the
chemiluminescence reaction between NO and O

3

in the reaction cell, to produce a

current signal proportional to the intensity of the chemiluminescence, to control the
temperature of the PMT to ensure the accuracy and stability of the measurements and to
drive the high voltage power supply that is needed for the PMT. The individual
functions are described individually below, Section 7.6.5 shows the sensor assembly and
its components.

8.5.2.1. Reaction Cell Heating Circuit

The stability of the chemiluminescence reaction between NO and O

3

can be affected by

changes in the temperature and pressure of the O

3

and sample gases in the reaction cell.

In order to reduce temperature effects, the reaction cell is maintained at a constant
50

 C, just above the high end of the instrument’s operation temperature range.

Two AC heaters, one embedded into the bottom of the reaction cell, the other embedded
directly above the chamber’s exhaust fitting, provide the heat source. These heaters
operate off of the instrument’s main AC power and are controlled by the CPU through a
power relay on the relay board (Section 8.5.7). A thermistor, also embedded in the
bottom of the reaction cell, reports the cell’s temperature to the CPU through the
thermistor interface circuitry of the motherboard (Section 8.5.9.3).

Advertising