Editing das parameters – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 90

Advertising
background image

Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

70

4.7.2.4. Editing DAS Parameters

Data channels can be edited individually from the front panel without affecting other
data channels. However, when editing a data channel, such as during adding, deleting or
editing parameters, all data for that particular channel will be lost, because the DAS can
store only data of one format (number of parameter columns etc.) for any given channel.
In addition, an DAS configuration can only be uploaded remotely as an entire set of
channels. Hence, remote update of the DAS will always delete all current channels and
stored data.

To modify, add or delete a parameter, follow the instruction shown in section 4.7.2.2
then press:


Edit Data Parameter Menu

Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 8, 800

PREV NEXT INS DEL

EDIT

PRNT

EXIT

From the DATA ACQUISITION menu

(see Section 6.7.2.2)

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC

<SET

SET>

EDIT PRINT

EXIT

SETUP X.X

PARAMETERS:

8

<SET SET>

EDIT

PRINT

EXIT

SETUP X.X

0) PARAM=DETREP, MODE=INST

PREV NEXT INS DEL EDIT EXIT

Press SET> key until…

SETUP X.X

EDIT PARAMS (DELETE DATA)


YES NO

NO returns to

the previous

menu and

retains all data.

Moves the

display between

available

Parameters

Inserts a new Parameter

before the currently

displayed Parameter

Deletes the Parameter

currently displayed.

Use to configure

the functions for

this Parameter.

Exits to the main

Data Acquisition

menu

YES will delete

all data in that

entire channel.

Advertising