Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 75

Advertising
background image

Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

55

in one of the DAS channels (Section 4.7) for data analysis or output on one of the
configurable analog outputs.

Table 4-2: Test Functions Defined

DISPLAY PARAMETER UNITS

DESCRIPTION

Analog output

range

configuration

A1:NXCNC1=100 PPM

A2:N0CNC1=100 PPM

A3:N2CNC1=25 PPM

A4:NXCNC2=100%

These functions show the default settings for the enabled analog
output channels. See section 4.13.4 for more information.

NOX STB

STABILITY PPM,

MGM

The stability is a standard deviation of the NO

X

concentration over 25

samples, each recorded every 10 seconds. A low NOX STB value
indicates low variability in NO

X

.

SAMP FLW

SAMPLE FLOW

cm³/min (cc/m)

The flow rate of the sample gas through the reaction cell. This value is
not measured but calculated from the sample pressure.

OZONE FL

OZONE

cm³/min (cc/m) Flow rate of the O

3

gas stream as measured with a flow meter

PMT

PMT Signal

MV

The raw output voltage of the PMT.

NORM PMT

NORMALIZED PMT

Signal

MV

The output voltage of the PMT after normalization for auto-zero offset and
temperature/pressure compensation (if activated).

AZERO

AUTO-ZERO MV

The PMT signal with zero NO

X

,

which is usually slightly different from 0 V.

This offset is subtracted from the PMT signal and adjusts for variations in
the zero signal.

HVPS

HVPS

V

The PMT high voltage power supply.

RCELL TEMP

REACTION CELL TEMP

C

The current temperature of the reaction cell.

BOX TEMP

BOX TEMPERATURE

C

The ambient temperature of the inside of the analyzer case.

PMT TEMP

PMT TEMPERATURE

C

The current temperature of the PMT.

CONV TEMP

CONVERTER

TEMPERATURE

C

The current temperature of the NO

2

converter.

RCEL

REACTION CELL

PRESSURE

in-Hg-A

The current gas pressure of the reaction cell as measured at the vacuum
manifold. This is the vacuum pressure created by the external pump.

SAMP

SAMPLE PRESSURE

in-Hg-A

The current pressure of the sample gas as it enters the reaction cell,
measured between the NO/NO

x

and Auto-Zero valves.

NOX SLOPE

NO

x

SLOPE

- -

The slope of the current NO

x

calibration as calculated from a linear fit

during the analyzer’s last zero/span calibration.

NOX OFFS

NO

x

OFFSET

MV

The offset of the current NO

x

calibration as calculated from a linear fit

during the analyzer’s last zero/span calibration.

NO SLOPE

NO SLOPE

- -

The slope of the current NO calibration as calculated from a linear fit
during the analyzer’s last zero/span calibration.

NO OFFS

NO OFFSET

MV

The offset of the current NO calibration as calculated from a linear fit
during the analyzer’s last zero/span calibration.

NO2

NO

2

concentration

PPM, MGM

The current NO

2

concentration in the chosen unit.

NOX

NO

x

concentration

PPM, MGM

The current NO

x

concentration in the chosen unit.

NO

NO concentration

PPM, MGM

The current NO concentration in the chosen unit.

TEST

TEST SIGNAL

2

MV

Signal of a user-defined test function on output channel A4.

TIME

CLOCK TIME

hh:mm:ss

The current day time for DAS records and calibration events.