Operating Instructions
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
127
To set the reporting range for an analog output, press:
EXAMPLE
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL
<SET SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 SCALE: 100.00 PPM
<SET SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 OUTPUT: 5V
SET>
EDIT
EXIT
DIAG AIO DATA_OUT_2 SCALE: 1250.00 PPM
<SET SET>
EDIT
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Use these
buttons to
change the
range scale.
Continue pressing SET> until ...
Continue pressing SET> until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
DIAG AIO DATA_OUT_2 SCALE: [1]00.00 PPM
<CH
CH>
INS
DEL
[1]
ENTR EXIT
DIAG AIO DATA_OUT_2 SCALE: 12[5]0. PPM
<CH
CH>
INS
DEL
[1]
ENTR
EXIT
RANGE SELECTION TOUCH SCREEN CONTROL BUTTON FUNCTIONS
BUTTON FUNCTION
<CH
Moves the cursor one character to the left.
CH>
Moves the cursor one character to the right.
INS
Inserts a character before the cursor location.
DEL
Deletes a character at the cursor location.
[?]
Press this key to cycle through the range of numerals and characters available for insertion:
0-9; as well as “+” & “-“.
ENTR
Accepts the new setting and returns to the previous menu.
EXIT
Ignores the new setting and returns to the previous menu.
Some keys only appear as needed.