Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

127

To set the reporting range for an analog output, press:

EXAMPLE

From the

AIO CONFIGURATION SUBMENU

(See Section 6.13.4.1)

DIAG AIO DATA_OUT_2 5V, NXCNC1, NOCAL

<SET SET>

EDIT

EXIT

DIAG AIO DATA_OUT_2 SCALE: 100.00 PPM

<SET SET>

EDIT

EXIT

DIAG AIO DATA_OUT_2 OUTPUT: 5V

SET>

EDIT

EXIT

DIAG AIO DATA_OUT_2 SCALE: 1250.00 PPM

<SET SET>

EDIT

EXIT

DIAG

ANALOG I/O CONFIGURATION

PREV NEXT

ENTR

EXIT

Use these

buttons to

change the

range scale.

Continue pressing SET> until ...

Continue pressing SET> until you reach the

output to be configured

DIAG AIO AOUTS CALIBRATED: NO

SET>

CAL

EXIT

DIAG AIO DATA_OUT_2 SCALE: [1]00.00 PPM

<CH

CH>

INS

DEL

[1]

ENTR EXIT

DIAG AIO DATA_OUT_2 SCALE: 12[5]0. PPM

<CH

CH>

INS

DEL

[1]

ENTR

EXIT

RANGE SELECTION TOUCH SCREEN CONTROL BUTTON FUNCTIONS

BUTTON FUNCTION

<CH

Moves the cursor one character to the left.

CH>

Moves the cursor one character to the right.

INS

Inserts a character before the cursor location.

DEL

Deletes a character at the cursor location.

[?]

Press this key to cycle through the range of numerals and characters available for insertion:

0-9; as well as “+” & “-“.

ENTR

Accepts the new setting and returns to the previous menu.

EXIT

Ignores the new setting and returns to the previous menu.

Some keys only appear as needed.