Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 251

Advertising
background image

Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

231

7.5.4.2. AC Configuration – Standard Heaters (JP2)

Power configuration for the AC the standard heaters is set using Jumper set JP2 (see
Figure 7-4 for the location of JP2).

Table 7-4: Power Configuration for Standard AC Heaters (JP2)

LINE VOLTAGE

JUMPER

COLOR

HEATER(S)

JUMPER

BETWEEN

PINS

FUNCTION

110 VAC / 115 VAC

50Hz & 60 Hz

WHITE

Reaction Cell / Sample

Chamber Heaters

1 to 8

Common

2 to 7

Neutral to Load

Mini Hi-Con

Converter

3 to 10

Common

4 to 9

Neutral to Load

Moly Converter

3 to 10

Common

4 to 9

Neutral to Load

Bypass Manifold

1

5 to 12

Common

6 to 11

Neutral to Load

220 VAC / 240 VAC

50Hz & 60 Hz

BLUE

Reaction Cell / Sample

Chamber Heaters

1 to 7

Load

Hi Concentration

Converter

3 to 9

Load

Moly Converter

3 to 9

Load

Bypass Manifold

1

5 to 11

Load

1

Bypass manifold is built into the reaction cell

1

2

3

4

5

6

7

8

9

10

11

12

Reaction Cell or

Sample Chamber

Heaters

Mini Hi-Con or

Moly Converter

Heaters

T200M/H

Bypass Manifold

Heater

110 VAC /115 VAC

1

2

3

4

5

6

7

8

9

10

11

12

Reaction Cell or

Sample Chamber

Heaters

Mini Hi-Con or

Moly Converter

Heaters

T200H/M

Bypass Manifold

Heater

220 VAC / 240 VAC

Figure 7-13:

Typical Set Up of AC Heater Jumper Set (JP2)