Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 305

Advertising
background image

Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

285

be adjusted through a potentiometer. These adjustments should only be performed when
encountering problems with the software calibration that cannot be rectified otherwise.
See Section 11.6.3.8 for this hardware calibration.

To

Motherboard

PMT Preamp PCA

Low

Pass Noise

Filter

E Test Control

From CPU

MUX

Amp to

Voltage

Converter/

Amplifier

D-A

Converter

PMT

Coarse

Gain Set

(Rotary

PMT Fine

Gain Set

(Rotary

Switch)

PMT

Signal

Offset

E-Test

Generator

O-Test

Generator

O Test Control

From CPU

PMT Temp

Sensor

PMT

Temperature

Feedback

Circuit

TEC Control

PCA

PMT Output

PMT HVPS

Drive Voltage

PMT Temp Analog Signal

to Motherboard

PMT Output Signal

(PMT) to Motherboard

O Test

LED

Figure 8-20:

PMT Preamp Block Diagram

The PMT temperature control loop maintains the PMT temperature around 7° C and can
be viewed as test function PMT TEMP on the front panel (see Section 6.2.1).

The electrical test (ETEST) circuit generates a constant, electronic signal intended to
simulate the output of the PMT (after conversion from current to voltage). By bypassing
the detector’s actual signal, it is possible to test most of the signal handling and
conditioning circuitry on the PMT preamplifier board. See section 6.9.6 for instructions
on performing this test.

The optical test (OTEST) feature causes an LED inside the PMT cold block to create a
light signal that can be measured with the PMT. If zero air is supplied to the analyzer,
the entire measurement capability of the sensor module can be tested including the PMT
and the current to voltage conversion circuit on the PMT preamplifier board. See
section 6.9.5 for instructions on performing this test.

Advertising