Set expected no and no, And vice versa) – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 195

Advertising
background image

Calibration Procedures

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

175

STEP TWO:

Set Expected NO and NO

X

Span Gas Concentrations.

These should be 80% of range of concentration values likely to be encountered in this
application. The default factory setting is 100 ppm. If one of the configurable analog
outputs is to be set to transmit concentration values, use 80% of the reporting range set
for that output (see Section 4.13.4)

SAMPLE

GAS TO CAL:NOX

NOX

O2

ENTR

EXIT

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST >

CAL

SETUP

M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX

<TST TST> ZERO SPAN

CONC

EXIT

SAMPLE

RANGE TO CAL:LOW

LOW

HIGH

ENTR

EXIT

M-P CAL CONCENTRATION MENU

NOX

NO

CONV

EXIT

M-P CAL NOX SPAN CONC:80.0 Conc

0

0

8

0

.0

ENTR EXIT

The NO

X

& NO span concentration

values automatically default to

80.0 Conc.

If this is not the the concentration of

the span gas being used, toggle

these buttons to set the correct

concentration of the NO

X

and NO

calibration gases.

EXIT ignores the new

setting and returns to

the previous display.

ENTR accepts the new

setting and returns to

the

CONCENTRATION

MENU.

If using NO span gas

in addition to NO

X

repeat last step.

Note

The expected concentrations for both NO

X

and NO are usually set to the

same value unless the conversion efficiency is not equal to 1.000 or not
entered properly in the conversion efficiency setting. When setting
expected concentration values, consider impurities in your span gas
source (NO often contains 1-3% NO

2

and vice versa).