Sample mode, Test functions – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 74

Advertising
background image

Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

54

The mode field of the front panel display indicates to the user which operating mode the
unit is currently running.

In addition to SAMPLE and SETUP, other modes the analyzer can be operated in are:

Table 4-1: Analyzer Operating modes

MODE EXPLANATION

SAMPLE

Sampling normally, flashing text indicates adaptive filter is on.

M-P CAL

This is the basic calibration mode of the instrument and is activated
by pressing the CAL key.

SETUP X.#

2

SETUP mode is being used to configure the analyzer. The gas
measurement will continue during this process.

SAMPLE A

Indicates that unit is in SAMPLE mode and AUTOCAL feature is
activated.

ZERO CAL M

1

Unit is performing ZERO calibration procedure initiated manually by
the user.

ZERO CAL A

1

Unit is performing ZERO calibration procedure initiated automatically
by the AUTOCAL feature.

ZERO CAL R

1

Unit is performing ZERO calibration procedure initiated remotely
through the COM ports or digital control inputs.

LO CAL A

Unit is performing LOW SPAN (midpoint) calibration initiated
automatically by the analyzer’s AUTOCAL feature.

LO CAL R

Unit is performing LOW SPAN (midpoint) calibration initiated remotely
through the COM ports or digital control inputs.

SPAN CAL M

1

Unit is performing SPAN calibration initiated manually by the user.

SPAN CAL A

1

Unit is performing SPAN calibration initiated automatically by the
analyzer’s AUTOCAL feature.

SPAN CAL R

1

Unit is performing SPAN calibration initiated remotely through the
COM ports or digital control inputs.

DIAG

One of the analyzer’s diagnostic modes is active (Section 4.13).

1

Only Appears on units with Z/S valve or IZS options.

2

The revision of the analyzer firmware is displayed following the word SETUP, e.g., SETUP

F.0.

The very important CAL mode, which allows calibration of the analyzer in various
ways, is described in detail in Section 7.

4.2. SAMPLE MODE

This is the analyzer’s standard operating mode. In this mode, the instrument is
analyzing NO and NO

X

and calculating NO

2

concentrations.

4.2.1. TEST FUNCTIONS

A series of test functions is available at the front panel while the analyzer is in SAMPLE
mode. These parameters provide information about the present operating status of the
instrument and are useful during troubleshooting (Section 7). They can also be recorded