4.13.7.2. Optic Test

The optic test function tests the response of the PMT sensor by turning on an LED
located in the cooling block of the PMT (Fig. 10-15). The analyzer uses the light
emitted from the LED to test its photo-electronic subsystem, including the PMT and the
current to voltage converter on the pre-amplifier board. To make sure that the analyzer
measures only the light coming from the LED, the analyzer should be supplied with zero
air. The optic test should produce a PMT signal of about 2000±1000 mV. To activate
the electrical test press the following touch screen button sequence.

SAMPLE RANGE = 500.0 PPB NOX=X.X


< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

ALRM

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR EXIT

DIAG OPTIC

A1:NXCNC1=100PPM NOX=XXX.X

<TST TST>

EXIT

SETUP X.X

ENTER DIAG PASS: 818

8 1 8

ENTR

EXIT

Press NEXT until…

DIAG

OPTIC TEST


PREV NEXT

ENTR

EXIT

Press TST until…

DIAG ELEC

PMT = 2751 MV

NOX=X.X


<TST TST>

EXIT

While the optic test is

activated, PMT should be

2000 mV ± 1000 mV

Note

This is a coarse test for functionality and not an accurate calibration tool.
The resulting PMT signal can vary significantly over time and also
changes with low-level calibration.