150

Table 4-29:

Status Output Pin Assignments

CONNECTOR

PIN

STATUS CONDITION

(ON=CONDUCTING)

1

SYSTEM OK

ON if no faults are present.

2

CONC VALID

ON if concentration measurement is valid, OFF when invalid.

3

HIGH RANGE

ON if unit is in high range of any AUTO range mode.

4

ZERO CAL

ON whenever the instrument is in ZERO calibration mode.

5

SPAN CAL

ON whenever the instrument is in SPAN calibration mode.

6

DIAG MODE

ON whenever the instrument is in DIAGNOSTIC mode.

7

LOW RANGE

ON if unit is in low range of any AUTO range mode.

8

Unused.

D

EMITTER BUS

The emitters of the transistors on pins 1-8 are bussed together. For
most applications, this pin should be connected to the circuit ground
of the receiving device.

+

DC POWER

+ 5 VDC, 30 mA maximum (combined rating with Control Inputs).

DIGITAL

GROUND

The ground from the analyzer’s internal, 5 VDC power supply.

4.15.1.2. Control Inputs

Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes
are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear
panel. These are opto-isolated, digital inputs that are activated when a 5 VDC signal
from the “U” pin is connected to the respective input pin.

Table 4-30: Control Input Pin Assignments

INPUT STATUS

CONDITION

WHEN

ENABLED

A

EXTERNAL ZERO

CAL

Zero calibration mode is activated. The mode field of the display
will read ZERO CAL R.

B

EXTERNAL SPAN

CAL

Span calibration mode is activated. The mode field of the display
will read SPAN CAL R.

C

EXTERNAL LOW

SPAN CAL

Low span (mid-point) calibration mode is activated. The mode field
of the display will read LO CAL R.

D, E & F

Unused

DIGITAL GROUND

Provided to ground an external device (e.g., recorder).

U

DC power for Input

pull ups

Input for +5 VDC required to activate inputs A - F. This voltage can
be taken from an external source or from the “+” pin.

+

Internal +5V Supply

Internal source of +5V which can be used to activate inputs when
connected to pin U.