Getting Started

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

25

Even though NO gas in nitrogen could be used as a span gas, the matrix of the balance
gas is different and may cause interference problems or yield incorrect calibrations. The
same applies to gases that contain high concentrations of other compounds (for example,
CO

2

or H

2

O). The span gas should match all concentrations of all gases of the measured

medium as closely as possible.

Cylinders of calibrated NO gas traceable to NIST-standard reference materials
specifications (also referred to as EPA protocol calibration gases) are commercially
available.

Table 3-9: NIST-SRM's Available for Traceability of NO

x

Calibration Gases

NIST-SRM

4

TYPE

NOMINAL

CONCENTRATION

2627a
2628a
2629a

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

5 ppm

10 ppm
20 ppm

1683b
1684b
1685b
1686b
1687b

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

50 ppm

100 ppm
250 ppm

5000 ppm
1000 ppm

2630

2631a

2635

2636a

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

Nitric Oxide (NO) in N

2

1500 ppm
3000 ppm

800 ppm

2000 ppm

2631a
1684b

Oxides of Nitrogen (NO

x

) in N

2

Oxides of Nitrogen (NO

x

) in N

2

3000 ppm

100 ppm

Note

If a dynamic dilution system such as the TAI Model T700 is used to dilute
high concentration gas standards to low, ambient concentrations, make
sure that the NO concentration of the reference gas matches the dilution
range of the calibrator. Choose an NO gas concentration that is in the
middle of the dilution system’s range.