Thermocouple inputs and configuration jumper (jp5) – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 307

Advertising
background image

Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

287

Thermistor(s) –

Low Temperature Sensing:

(e.g. Sample Chamber and Reaction
Cell temperatures)

RELAY PCA

DC

Control

Logic

Solid State

AC Relays

Preamplifiers

and Signal

Conditioning

MOTHER BOARD

A/D

Converter

(V/F)

CPU

Themocouple(s)

(High Temperature Sensing;

e.g. Moly and HiCon

Converter temperatures)

AC HEATERS

DC HEATERS

THERMOCOUPLE
CONFIGURATION

JUMPER

(JP5)

Cold Junction

Compensation

Figure 8-21:

Heater Control Loop Block Diagram.

8.5.7.3. Thermocouple Inputs and Configuration Jumper (JP5)

Although the relay PCA supports two thermocouple inputs, the current 9110TH/M
series analyzers only utilize one. By default, this single thermocouple input is plugged
into the TC1 input (J15). TC2 (J16) is currently not used. See Figure 7-4 for location of
J15 and J16

CAUTION

Avoid damage to the unit: use only the recommended thermocouple type and its specific

settings. If in doubt, call T-API Technical Support for information about the correct part.