Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

57

Table 4-3: List of Warning Messages

MESSAGE DEFINITION

ANALOG CAL WARNING

The instrument’s analog-to-digital converter (A/D) circuitry or one of the analog
outputs are not calibrated.

AZERO WRN XXX.X MV

The reading taken during the Auto-zero cycle is outside the specified limits. The
value shown here as “XXX.X” indicates the actual auto-zero reading at the time of
the warning.

BOX TEMP WARNING

The temperature inside the 9110TH/M chassis is outside the specified limits.

CANNOT DYN SPAN

Remote span calibration failed while the dynamic span feature was ON.

CANNOT DYN ZERO

Remote zero calibration failed while the dynamic zero feature was ON.

CONFIG INITIALIZED

Configuration storage was reset to factory configuration or was erased.

CONV TEMP WARNING

NO

2

converter temperature is outside of specified limits.

DATA INITIALIZED

DAS data storage was erased.

HVPS WARNING

High voltage power supply for the PMT is outside of specified limits.

OZONE FLOW WARNING

Ozone flow is outside of specified limits.

OZONE GEN OFF

Ozone generator is off. This is the only warning message that automatically
clears itself when the ozone generator is turned on.

PMT TEMP WARNING

PMT temperature is outside of specified limits.

RCELL PRESS WARN

Reaction cell pressure is outside of specified limits.

RCELL TEMP WARNING

Reaction cell temperature is outside of specified limits.

REAR BOARD NOT DET

The firmware is unable to communicate with the motherboard.

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SYSTEM RESET

The computer rebooted or was powered up.

To view and clear warning messages

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

TEST

CAL

MSG

CLR

SETUP

Make sure warning messages are

not due to real problems.

Press CLR to clear the current

message.

If more than one warning is active, the

next message will take its place
Once the last warning has been

cleared, the analyzer returns to

SAMPLE mode

SAMPLE

A1:NXCNC1=100PPM

NO=XXX.X


< TST TST > CAL

MSG

CLR

SETUP

SAMPLE

HVPS WARNING

NO2=XXX.X


TEST CAL

MSG

CLR

SETUP

TEST deactivates warning

messages

MSG activates warning

messages.

<TST TST> keys replaced with

TEST key

All Warning messages are hidden,

but MSG button appears

NOTE:

If the warning message persists

after several attempts to clear it,

the message may indicate a

real problem and not an artifact

of the warm-up period

Figure 4-3:

Viewing and Clearing 9110TH/M WARNING Messages