Set the expected no, Span gas concentration, Step two – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 190

Advertising
background image

Model 9110TH NOx Analyzer

Calibration Procedures

Teledyne Analytical Instruments

170

STEP TWO:

Set the expected NO

2

span gas concentration:

SAMPLE

GAS TO CAL:NOX

NOX

O2

ENTR

EXIT

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST >

CAL

SETUP

M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX

<TST TST> ZERO SPAN

CONC

EXIT

SAMPLE

RANGE TO CAL:LOW

LOW

HIGH

ENTR

EXIT

M-P CAL CONCENTRATION MENU

NOX

NO

CONV

EXIT

M-P CAL NO2 CE CONC:80.0 Conc

0

0

8

0

.0

ENTR EXIT

The NO

X

& NO span concentration

values automatically default to

80.0 Conc.

If this is not the the concentration of

the span gas being used, toggle

these buttons to set the correct

concentration of the NO

X

and NO

calibration gases.

EXIT ignores the new

setting and returns to

the previous display.

ENTR accepts the new

setting and returns to

the

CONVERTER

EFFICIENCY

MENU.

If using NO span gas

in addition to NO

X

repeat last step.

M-P CAL CONVERTER EFFICIENCY MENU

NO2

CAL

SET

EXIT