Step four – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 192

Advertising
background image

Model 9110TH NOx Analyzer

Calibration Procedures

Teledyne Analytical Instruments

172

STEP FOUR:

Perform the converter efficiency calculation procedure:

SAM PLE

NO 2 STB= XXX.X PPM

NO X =XXX.X

< TST TST >

CAL

SETU P

SAM PLE

G AS TO CAL:NO X

NO X

O 2

ENTR

EXIT

M -P C AL STB = XXX.X PPM O X =X.XXX

<TST TST> ZER O SPAN

C O NC

EXIT

SAM PLE

RANG E TO CAL:LO W

LO W

HIG H

ENTR

EXIT

M -P C AL

CO N CENTRATIO N M EN U

N O X

NO

C O NV

EXIT

M -P C AL

C E FACTO R :1.000 G ain

1

.0

0

0

0

ENTR

EXIT

Press EXIT 3 tim es

top return to the

SAM PLE display

M -P C AL

C O NVER TER EFFICIEN CY M ENU

NO 2

C AL

SET

EXIT

SAM PLE A 1:NXCNC 1=100PPM NO X =XXX.X

< TST

TST >

CAL SETU P

Set the D isplay to show

the N O 2 STB test

function.

This function calculates

the stability of the

m easurem ent

Toggle TST> button until ...

Allow NO

2

to enter the sam ple port

at the rear of the analyzer.

W ait until NO 2 STB

falls below 0.5 ppm

and the ENTR button

appears.

This m ay take several

m inutes.

SAM PLE

NO X STB= XXX.X PPM

N O X =XXX.X

< TST TST >

ENTR

SETU P

M -P C AL

C O NVER TER EFFICIEN CY M ENU

NO 2

C AL

SET

EXIT

M -P C AL

C E FACTO R :1.012 G ain

1

.0

0

1

2

ENTR

EXIT

W hen ENTR is

pressed, the ratio of

observed NO

2

concentration to

expected NO

2

concentration is

calculated and

stored.

M -P C AL

C O NVER TER EFFICIEN CY M ENU

NO 2

C AL

SET

EXIT