Operation within the 9110th/m analyzer, Pneumatic operation of the o, Sensor – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 298

Advertising
background image

Model 9110TH NOx Analyzer

Principles

of

Operation

Teledyne Analytical Instruments

278

Figure 8-14:

Oxygen Sensor - Principle of Operation

8.4.2. OPERATION WITHIN THE 9110TH/M ANALYZER

The oxygen sensor option is transparently integrated into the core analyzer operation.
All functions can be viewed or accessed through the front panel, just like the functions
for NO

X

.

 The O2 concentration is displayed in the upper right-hand corner,

alternating with NOX, NO and NO2 concentrations.

 Test functions for O2 slope and offset are viewable from the front panel

along with the analyzer’s other test functions.

 O2 sensor calibration is performed via the front panel CAL function and is

performed in a nearly identical manner as the standard NOX/NO
calibration. See Section 5 for more details.

 Stability of the O

2

sensor can be viewed (see 3.3.2.1)

The O

2

concentration range is 0-100% (user selectable) with 0.1% precision and

accuracy and is available to be output via one of the instrument’s four user selectable
analog outputs (see Section 6.13.4).

The temperature of the O

2

sensor is maintained at a constant 50° C by means of a PID

loop and can be viewed on the front panel as test function O2 TEMP.

The O

2

sensor assembly itself does not have any serviceable parts and is enclosed in an

insulated canister.

8.4.3. PNEUMATIC OPERATION OF THE O

2

SENSOR

Pneumatically, the O

2

sensor is connected after the particulate filter and draws a flow of

about 80 cm³/min in addition to the normal sample flow rate (See Table 10.-3 for
nominal sample inlet gas flow rates) and is separately controlled with its own critical
flow orifice located inside the vacuum manifold.

Advertising