Ain calibration, Gain (number of units represented by 1 volt) – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 158

Advertising
background image

Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

138

4.13.6.4. AIN Calibration

This is the sub-menu calibrates the analyzer’s A-to-D conversion circuitry. This
calibration should only be necessary after major repair such as a replacement of CPU,
motherboard or power supplies.

To perform a AIN CALIBRATION, press:

From the

AIO CONFIGURATION SUBMENU

(See Section 6.13.4.1)

DIAG

ANALOG I/O CONFIGURATION

PREV NEXT

ENTR

EXIT

Continue pressing SET> until ….

DIAG AIO AIN CALIBRATED: NO

<SET

CAL

EXIT

DIAG AIO CALIBRATING A/D ZERO

DIAG AIO CALIBRATING A/D SPAN

DIAG AIO AIN CALIBRATED: YES

<SET

CAL

EXIT

DIAG AIO AOUTS CALIBRATED: NO

SET>

CAL

EXIT

4.13.6.5. Configuring External Analog Inputs (Option) Channels

To configure the analyzer’s external analog inputs option, define for each channel:

 gain (number of units represented by 1 volt)