4.13.7.3. Electrical Test

The electrical test function creates a current, which substitutes the PMT signal, and
feeds it into the preamplifier board. This signal is generated by circuitry on the pre-
amplifier board itself and tests the filtering and amplification functions of that assembly
along with the A/D converter on the motherboard. It does not test the PMT itself. The
electrical test should produce a PMT signal of about 2000 ±1000 mV.

To activate the electrical test press the following buttons:

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR EXIT

DIAG ELEC

A1:NXCNC1=100PPM NOX=XXX.X


<TST TST>

EXIT

Press NEXT until…

DIAG

ELECTRICAL TEST

PREV NEXT

ENTR

EXIT

Press TST until…

DIAG ELEC

PMT = 1732 MV

NOX=X.X

<TST TST>

EXIT

While the electrical test is

activated, PMT should equal:

2000 mV ± 1000 mV

SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR

EXIT