Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 165

Advertising
background image

Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

145

4.13.7.4. Ozone Generator Override

This feature allows the user to manually turn the ozone generator off and on. This can
be done before disconnecting the generator, to prevent ozone from leaking out, or after a
system restart if the user does not want to wait for 30 minutes during warm-up time.
Note that this is one of the two settings in the DIAG menu that is retained after you exit
the menu. (During initial power up TMR (timer) is displayed while the Ozone brick
remains off for the first 30 minutes). Also note that the ozone generator does not turn on
if the ozone flow conditions are out of specification (e.g., if there is no flow through the
system or the pump is broken).

Advertising