Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

76

4.7.2.8. RS-232 Report Function

The 9110TH/M DAS can automatically report data to the communications ports, where
they can be captured with a terminal emulation program or simply viewed by the user.

To enable automatic COM port reporting, follow the instruction shown in section 4.7.2.2
then press:

Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 8, 800

PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC

<SET

SET>

EDIT PRINT

EXIT

SETUP X.X

RS-232 REPORT: OFF

<SET SET>

EDIT

PRINT

EXIT

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

menu.

SETUP X.X

RS-232 REPORT: OFF

OFF

ENTR EXIT

Toggle button to turn

reporting ON or OFF

Press SET> key until…

From the DATA ACQUISITION menu

(see Section 6.7.2.2)

4.7.2.9. Compact Report

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead
of reporting each parameter in one channel on a separate line, up to five parameters are
reported in one line, instead. For example, channel DIAG would report its record in two
lines (10 parameters) instead of 10 lines. Individual lines carry the same time stamp and
are labeled in sequence.

4.7.2.10. Starting Date

This option allows to specify a starting date for any given channel in case the user wants
to start data acquisition only after a certain time and date. If the Starting Date is in the
past, the DAS ignores this setting.

Advertising