Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 132
Model 9110TH NOx Analyzer
Operating
Instructions
Teledyne Analytical Instruments
112
To enter the signal I/O test mode, press:
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
EXAMPLE
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
DIAG I / O
Test Signals Displayed Here
P
REV NEXT
JUMP
PRNT EXIT
Press JUMP to go
directly to a specific
signal
See Appendix A-4 for
a complete list of
available SIGNALS
DIAG I / O
JUMP TO: 5
0 5
ENTR
EXIT
Enter 05 to Jump
to Signal 5:
(CAL_LED)
DIAG I / O
CAL_LED = ON
PREV NEXT JUMP
ON
PRNT EXIT
Exit to return
to the
DIAG menu
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
Use the NEXT & PREV
keys to move between
signal types.
Pressing the PRNT key will send a formatted printout to the serial port and can be
captured with a computer or other output device.
EXIT
returns
to the main
SAMPLE
display