background image

Getting Started

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

39

3.5.3.3. Internal Flow for O

2

Sensor Option 65A

Please see Section 3.7.2 for calibration connections and method.

VACUUM

PRESSURE

SENSOR

SAMPLE

PRESSURE

SENSOR

O3

F

L

OW

SE

N

S

OR

FLOW PRESSURE

SENSOR PCA

INSTRUMENT CHASSIS

NO/NO

X

VALVE

AUTOZERO

VALVE

O

3

Destruct

O

3

GENERATOR

EXHAUST

GAS

OUTLET

SAMPLE

GAS

INLET

E

X

HA

UST

MANIF

O

LD

O

2

Sensor

Orifice Dia.

0.007"

Orifice Dia.

0.004"

PUMP

NO

X

Exh

aus

t

S

cr

ubb

er

Orifice Dia.

0.007"

REACTION

CELL

PMT

PERMAPURE
DRYER

Fi

lt

e

r

Orifice Dia.

0.004"

OPTION, O2

SENSOR, P/N 04453

NC

COM

NO

NC

COM

NO

NO

2

Converter

Figure 3-23:

9110TH – Internal Pneumatics with O2 Sensor Option 65A