Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 53
Getting Started
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
33
Figure 3-20:
9110TM – Internal Pneumatics with Ambient Zero-Span Valve Option 50A
Table 3-10: Zero/Span Valve States
MODE VALVE
CONDITION
SAMPLE
Sample/Cal
Open to sample gas inlet
Zero/Span
Open to zero air inlet
ZERO
CALIBRATION
Sample/Cal
Open to zero/span inlet (activated)
Zero/Span
Open to zero air inlet
SPAN
CALIBRATION
Sample/Cal
Open to zero/span inlet (activated)
Zero/Span
Open to span gas inlet / IZS gas (activated)
The state of the zero/span valves can also be controlled:
Manually from the analyzer’s front panel by using the SIGNAL I/O
controls located under the DIAG Menu (Section 4.13.2),
By activating the instrument’s AutoCal feature (Section 5.8),
Remotely by using the external digital control inputs (Section 4.15.1.2) or
Ethernet option.
Remotely through the RS-232/485 serial I/O ports (Section 4.11).