Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 53

Advertising
background image

Getting Started

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

33

Figure 3-20:

9110TM – Internal Pneumatics with Ambient Zero-Span Valve Option 50A

Table 3-10: Zero/Span Valve States

MODE VALVE

CONDITION

SAMPLE

Sample/Cal

Open to sample gas inlet

Zero/Span

Open to zero air inlet

ZERO

CALIBRATION

Sample/Cal

Open to zero/span inlet (activated)

Zero/Span

Open to zero air inlet

SPAN

CALIBRATION

Sample/Cal

Open to zero/span inlet (activated)

Zero/Span

Open to span gas inlet / IZS gas (activated)

The state of the zero/span valves can also be controlled:

 Manually from the analyzer’s front panel by using the SIGNAL I/O

controls located under the DIAG Menu (Section 4.13.2),

 By activating the instrument’s AutoCal feature (Section 5.8),

 Remotely by using the external digital control inputs (Section 4.15.1.2) or

Ethernet option.

 Remotely through the RS-232/485 serial I/O ports (Section 4.11).