4.13.7. OTHER DIAG MENU FUNCTIONS

4.13.7.1. Display Sequence Configuration

The model 9110TH/M analyzer allows the user to choose which gas concentration
measurement and reporting range is to be displayed in the concentration field on the
instrument’s front panel display as well as what order and how long each will appear
before analyzer cycle to the next item on the display list.

Note

This 9110TH/M is constantly monitoring all of the gas measurements it is
configured to make regardless of which range is being displayed. This
feature merely changes how that display sequence occurs, not how the
instrument makes measurements.

The software permits the user to choose from the following list of display values:

Table 4-26: 9110TH/M Available Concentration Display Values

DISPLAY

VALUE

DESCRIPTION

ASSOCIATED DAS

PARAMETER

NOX

NO

x

value computed with the slope and offset values for the currently selected

NO

x

range.

1

--

NXL

NO

x

value computed with the slope and offset values for NO

x

reporting range

1 (Low)

NXCNC1

NXH

NO

x

value computed with the slope and offset values for NO

x

reporting range

2 (High)

NXCNC2

NO

NO value of computed with the slope and offset values for the currently
selected NO range

1

--

NOL

NO value computed with the slope and offset values for NO reporting range 1
(Low)

NOCNC1

NOH

NO value computed with the slope and offset values for NO reporting range 2
(High)

NOCNC2

N2

NO

2

value of computed with the slope and offset values for the currently

selected NO

2

range

1

--

N2L

NO

2

value computed for with the slope and offset values for NO

x

reporting

range 1 (Low) & N0 reporting range 1 (Low)

N2CNC1

N2H

NO

2

value computed for with the slope and offset values for NO

x

reporting

range 2 (High) & N0 reporting range 2 (High)

N2CNC2

O2

O

2

concentration value.

O2CONC

2

1

With the following exceptions this will be reporting range 1 (Low) for the appropriate gas type:

 If the analyzer is in calibration mode, this will be the concentration value computed with the slope and offset for which ever range

is being calibrated.

 If the instrument is in either E-Test or O-Test mode, this will be the value computed with the slope and offset values used by

those tests.

2

Only appears if O

2

sensor option is installed.