3.6.3. FUNCTIONAL CHECK

After the analyzer’s components have warmed up for at least 30 minutes, verify that the
software properly supports any hardware options that were installed.

Check to make sure that the analyzer is functioning within allowable operating
parameters. Appendices A and C include a list of test functions viewable from the
analyzer’s front panel as well as their expected values. These functions are also useful
tools for diagnosing performance problems with your analyzer (Section 7). The
enclosed Final Test and Validation Data Sheet (part number 04490) lists these values
before the instrument left the factory. To view the current values of these test functions
press the <TST TST> buttons:

A1:NXCNC1=100 PPM

1

A2:N0CNC1=100 PPM

1

A3:N2CNC1=25 PPM

1

A4:NXCNC2=100%

1

NOX STB

SAMP FLOW
OZONE FLOW

PMT

NORM PMT

AZERO

HVPS
RCELL TEMP

BOX TEMP

PMT TEMP

MF TEMP
O2 CELL TEMP

2

MOLY TEMP

RCEL

SAMP
NOX SLOPE

NOX OFFSET

NO SLOPE

NO OFFSET
O2 SLOPE

2

O2 OFFSET

2

TIME

SAMPLE

A1:NXCNC1=100 PPM NOX = XXX


< TST TST >

CAL

SETUP

1

default settings for user

selectable reporting range
settings.

2

Only appears if O

2

sensor

option is installed.

Toggle <TST TST> to scroll