Connecting the communications ports – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 42

Advertising
background image

Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

22

Table 3-7: Concentration Alarm Relay Output Operation

RELAY

FUNCTION

RELAY PIN

STATE

1

COMMENTS

N
O

C

N
C

AL2

Concentration Alarm 1

Active

Gas

concentration level is above the trigger limit set for

CONC_ALARM_1

 DAS Trigger CONCW1 ACTIVATED
CONC ALARM1 WARN appears on Analyzer Display

Concentration Alarm 1

Inactive

Gas

concentration level is below the trigger limit set for

CONC_ALARM_1

AL3

Concentration Alarm 2

Active

Gas

concentration level is above the trigger limit set for

CONC_ALARM_2

 DAS Trigger CONCW2 ACTIVATED
CONC ALARM2 WARN appears on Analyzer Display

Concentration Alarm 2

Inactive

Gas

concentration level is below the trigger limit set for

CONC_ALARM_2

1

NO = Normally Open operation.
C = Common
NC = Normally Closed operation.

3.4.8. CONNECTING THE COMMUNICATIONS PORTS

For RS-232 or RS-485 (option) communications through the analyzer’s two serial
interface ports, refer to Section 4.11 for information and connection instructions.

3.4.8.1. Connecting to a LAN or the Internet

For network or Internet communication with the analyzer, connect an Ethernet cable
from the analyzer’s rear panel Ethernet interface connector to an Ethernet port. See
Section 4.11.7 for configuration instructions.

Note

The 9110TH/M firmware supports dynamic IP addressing or DHCP. If your
network also supports DHCP, the analyzer will automatically configure its
LAN connection appropriately. If your network does not support DHCP,
see Section 4.11.7.2 for instructions on manually configuring the LAN
connection.

3.4.8.2. Connecting to a Personal Computer (PC)

If the USB port is configured for direct communication between the analyzer and a
desktop or a laptop PC, connect a USB cable between the analyzer and the PC or laptop
USB ports, and follow the set-up instructions in Section 4.11.8. (RS-485 communication
is not available with the USB com port option).