Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

134

4.13.6.2. Manual Calibration of Analog Output Configured for Voltage Ranges

For highest accuracy, the voltages of the analog outputs can be manually calibrated.

Note

The menu for manually adjusting the analog output signal level will only
appear if the AUTO-CAL feature is turned off for the channel being
adjusted.

Calibration is performed with a voltmeter connected across the output terminals (See
Figure 6-14) and by changing the actual output signal level using the front panel keys in
100, 10 or 1 count increments.

V

+DC Gnd

V OUT +

V OUT -

V IN +

V IN -

Recording
Device

ANALYZER

See the Electrical

Connections

section for pin

assignments of

Analog Out

connector on the

rear panel

Figure 4-11:

Setup for Calibrating Analog Outputs

Table 4-24: Voltage Tolerances for Analog Output Calibration

FULL

SCALE

ZERO

TOLERANCE

SPAN VOLTAGE

SPAN

TOLERANCE

MINIMUM

ADJUSTMENT

(1 count)

0.1 VDC

±0.0005V

90 mV

±0.001V

0.02 mV

1 VDC

±0.001V

900 mV

±0.001V

0.24 mV

5 VDC

±0.002V

4500 mV

±0.003V

1.22 mV

10 VDC

±0.004V

4500 mV

±0.006V

2.44 mV