2 live display - tracks, Live display - tracks – Metrohm viva 1.1 User Manual

Page 172

Advertising
background image

3.8 Live display subwindow

■■■■■■■■■■■■■■■■■■■■■■

160

■■■■■■■■

viva 1.1

Application note
Shows the application note defined in the START command.

Messages
Shows the messages generated during the determination.

3.8.2

Live display - Tracks

Subwindow: Workplace

Live display 1 / Live display 2

The main track of the loaded method (START command) is always shown
in the Live display subwindow. Further tracks are only displayed if the
live display is enabled in the corresponding TRACK command. A tab
labeled with the name of the track is displayed per track.

The live display elements (curve, measured value, message) defined in the
Properties are displayed for the currently active command on these tabs.
You have to switch manually between the tabs, i.e., there is no automatic
switching when another track is called. This makes it possible to display
two simultaneously active tracks next to each other in two live windows.

At the start of the determination the content of the track tabs is deleted.
Afterwards the messages, curves and measured value displays defined by
the active commands will appear in the track tabs. The tab titles of the
active tracks will be displayed in red. Additionally, the command type and
command name of the active command are displayed in the window title
of the Live display subwindow.

If all tracks or individual tracks are stopped with [Hold] or with a SEND
command, respectively, then Track halted... is displayed in these tracks.

After completion of the track, Track finished is displayed in all tracks
except for the main track. After the determination has been finished, the
following is displayed in the main track, depending on the determination
run:

Selection

Run: regular without remarks | Run: regular
with remarks
| Run: Stop | Run: Stop by error

Run: regular without remarks
The determination has been finished automatically after it was pro-
cessed regularly and without any remarks.
Run: regular with remarks
The determination has been finished automatically after it was pro-
cessed regularly but with remarks (see Chapter 4.7.6, page 365).
Run: Stop
The determination has been canceled either manually with [Stop], by
stop criteria or a SEND command.
Run: Stop by error
The determination has been canceled automatically as a result of an
error or not even started as a result of an error at the start test.

Advertising