6 properties - msb, Properties - msb – Metrohm viva 1.1 User Manual

Page 879

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Devices

viva 1.1

■■■■■■■■

867

for Move = Rack position
Input range

1 - 999

Default value

1

for Move = Special beaker
Input range

1 - 16

Default value

1

for Move = External position
Input range

1 - 4

Default value

1

7.4.6

Properties - MSB #

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

MSB #

The MSB # tabs show the properties of the MSB connector and the devi-
ces connected to it.

Request for dosing device preparation
Selects when to display the request to execute the PREP command (see
Chapter 5.5.2.4.7.1, page 467)
(prepare) for the dosing device connected
to the MSB.

At program start

on | off (Default value: on)

If this check box is activated, then the request to prepare the dosing
device is displayed every time the program is started.

When attaching a dosing unit

on | off (Default value: on)

If this check box is activated, then the request to prepare the dosing
device is displayed every time a dosing unit is mounted.

Time interval

on | off (Default value: off)

If this check box is activated, then the request to prepare the dosing
device is displayed after the time interval defined below.

all

Time interval after which the request to prepare the dosing device is dis-
played.

Input range

0.1 - 999.9 h

Default value

12 h