Page 484

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

472

■■■■■■■■

viva 1.1

Selection

846 Dosing Interface | 814 USB Sample Pro-
cessor
| 815 Robotic USB Sample Processor
XL
| 858.0010 Professional Sample Processor |
858.0020 Professional Sample Processor |
858.0030 Professional Sample Processor |
919.0020 IC Autosampler plus | 894 Profes-
sional CVS
| 089 Photometer

Default value

894 Professional CVS

Dosing device

Dosing device

Selection of the dosing device (Dosino only) to be emptied. All the dosing
device connections which are possible with the selected device type are
always displayed.

846, 894
Selection

1 | 2 | 3 | 4

Default value

1

814, 815, 858, 919
Selection

1 | 2 | 3

Default value

1

089
Selection

1

Default value

1

Name of dosing unit

Selection of a dosing unit from the dosing units available in the table of
the dosing units.

Selection

'Name of the dosing unit' | not defined

Default value

not defined

'Name of the dosing unit'
Alphabetically organized selection of the names of the dosing units
from the table of dosing units in the configuration. A check is made
prior to the start of a method to determine whether the selected dos-
ing unit is identical with the name of the dosing unit at the selected
MSB connector.
not defined
No check is made prior to the start of a method to determine whether
the selected dosing unit is identical with the name of the dosing unit at
the selected MSB connector. This has the advantage that a method can
be utilized by users who use other dosing units.