4 lat/mlat calibration curve properties – Metrohm viva 1.1 User Manual

Page 682

Advertising
background image

5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

670

■■■■■■■■

viva 1.1

5.6.5.4

LAT/MLAT calibration curve properties

Dialog window: Method

Evaluation - Calibration Calibration

curves

[Edit] Properties... Properties calibration

The calibration curve parameters for the substance selected in the table
can be edited in the Properties calibration dialog window.

Measuring command

Shows the selected measuring command.

Substance

Shows the selected substance.

Evaluation quantity

Selection of the quantity according to which the substance is to be evalu-
ated.

Selection

Area | Height

Default value

Area

Weighting

on | off (Default value: on)

If this check box is activated, then the points on the calibration curve will
be weighted in accordance with the formula 1/evaluation quantity

².

Curve type

Only the curve type Linear regression is permitted for the calibration
methods LAT and MLAT. At least two calibration points of different con-
centrations are required for determining the line.

Selection

Linear regression | Quadratic regression | Non-
linear regression
| Linear interpolation

Default value

Linear regression

Curve type

Calibration function

Linear regression

y = CALC 0 + CALC 1

x

Navigation bar

Navigation bar for changing the displayed substances with the following
functions:

Jump to the first substance. Altered data is adopted.