2 digital in - general – Metrohm viva 1.1 User Manual

Page 570

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

558

■■■■■■■■

viva 1.1

Mode of operation
At the time of the command, the statuses of all of the digital inputs are
entered and saved in command variables (syntax 'Command name.Vari-
able identification'
)

Parameters
The parameters for the DIGITAL IN command are set on the following
tabs:

General
Parameters for instrument.

Stop criteria
Parameters for the definition of the stop criteria.

Command variables
The following command variables are generated by the DIGITAL IN com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Port name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

[VAL]

Value of the command variables (optional, i.e. 'Com-
mand name.Portname'
= 'Command
name.Portname.VAL'
). The value returned corre-
sponds to the status of the IO port after the com-
mand is exited; 1 = active, 0 = inactive

5.5.2.8.3.2

DIGITAL IN - General

Tab: Method

DIGITAL IN Properties... DIGITAL IN - 'Command

name'

General

Command name

Name of the command.

Entry

25 characters

Advertising