Page 569
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1
■■■■■■■■
557
Parameters
Parameter 1
The meaning of Parameter 1 depends on the Type set for the digital
output.
For Type = Stirrer: stirring performance in %, pulse duration dependent
on Parameter 2.
Input range
-214,783,647 - 214,783,647
Default value
100
Parameter 2
The meaning of Parameter 2 depends on the Type set for the digital
output.
For Type = Stirrer:
■
0: Duration of the switch-on pulse set to 500 ms.
■
>0: With a stirring performance of 100% (Parameter 1 = 100), pulse
duration 10 ms
Example:
Parameter 1 = 100, Parameter 2 = 30: pulse duration switched on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: pulse duration switched on
225 ms, switched off 75 ms
Input range
-214,783,647 - 214,783,647
Default value
0
5.5.2.8.3
DIGITAL IN
5.5.2.8.3.1
DIGITAL IN - Overview
Dialog window: Method
▶ DIGITAL IN ▶ Properties... ▶ DIGITAL IN - 'Command
name'
Instruments
This command can be executed with the IO controller and the Stepping
Motor Controller.
Appearance
The command has the following appearance: