Metrohm viva 1.1 User Manual

Page 547

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

535

Input range

0 - 100 pixels

Default value

0 pixels

Tab for the 089 Photometer

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Input range

–1E+12 - 1E+12 (max. 10 digits)

Default value

0.0

Entry

1 - 10 characters

Selection

mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | µg/mL | ppm | % | mEq/L

Default value

ppm

Measurement
Two types of measurement can be selected, Measurement with drift
control
or Measurement without drift control.

Selection

Measurement with drift control | Measure-
ment without drift control

Default value

Measurement with drift control

Measurement with drift control
The measurement is carried out drift-controlled if this option is
selected. The measurement is canceled as soon as the defined Signal
drift
or the Stop measured value is reached or a defined Waiting
time
has elapsed. The corresponding parameters are not visible if this
option is disabled.
Measurement without drift control
The measurement is carried out without drift control if this option is
selected. Measurement is continued for as long as one of the two stop
criteria Max. waiting time or Stop measured value is fulfilled. The
corresponding parameters are not visible if this option is disabled.