2 digital out - general – Metrohm viva 1.1 User Manual

Page 565

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

553

The timer function can be used to set an output to active for a certain
time after a start delay.

After the command has run, the statuses and the default values of all of
the digital outputs used in the command are saved in command variables.

Parameters
The parameters for the DIGITAL OUT command are set on the following
tabs:

General

Digital outputs

Command variables
The following command variables are generated by the DIGITAL OUT
command in the method run and can be used in formulas under the des-
ignation 'Command name.Port name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

DEF

Default value which is defined in the configuration
for an IO port; 1 = active, 0 = inactive

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

[VAL]

Value of the command variables (optional, i.e. 'Com-
mand name.Portname'
= 'Command
name.Portname.VAL'
). The value returned corre-
sponds to the status of the IO port after the com-
mand is exited; 1 = active, 0 = inactive

5.5.2.8.2.2

DIGITAL OUT - General

Tab: Method

DIGITAL OUT Properties... DIGITAL OUT - 'Command

name'

General

Command name

Name of the command.

Entry

25 characters