5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

518

■■■■■■■■

viva 1.1

Identification

Description

BP{#}.[MEA]

Measured value for the break point (1 - 9) in the unit
of the measured value

BP{#}.CNT

Intensity for the break point x (1 - 9) in counts

BP{#}.TEM

1)

Temperature for the break point # (1 - 9) in °C

BP{#}.TIM

Time for the break point # (1 - 9) in s

BP{#}.TRN

Transmission for the break point x (1 - 9) in %

BSY

Command status:

1 = BUSY, HOLD or ERROR

0 = READY

invalid (variable not available) = Command has
never been started

DBL

Total duration for the processing of the command in
s

EME

End measured value (measured value after processing
of the command) in the unit of the measured value

ETE

1)

End temperature (temperature after the command
has been processed) in °C

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

invalid (variable not available) = Command has
never been started

FP{#}.[MEA]

Measured value for the fixed point # (1 - 9) in the unit
of the measured value

FP{#}.CNT

Intensity for the fixed point x (1 - 9) in counts

FP{#}.TEM

1)

Temperature for the fixed point # (1 - 9) in °C

FP{#}.TIM

Time for fixed point # (1 - 9) in s

FP{#}.TRN

Transmission for the fixed point x (1 - 9) in %

IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

ITE

1)

Initial temperature (temperature before start condi-
tions are processed) in °C