6 rc calibration curve properties – Metrohm viva 1.1 User Manual

Page 687

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

675

5.6.5.6

RC calibration curve properties

Dialog window: Method

Evaluation - Calibration Calibration

curves

[Edit] Properties... Properties calibration

The calibration curve parameters for the substance selected in the table
can be edited in the Properties calibration dialog window.

Measuring command

Shows the selected measuring command.

Substance

Shows the selected substance.

Evaluation quantity

Selection of the quantity according to which the substance is to be evalu-
ated.

Selection

Area | Height

Default value

Area

Area
For the measured quantity Current, the area is converted to the corre-
sponding charge.
Height
For the measured quantity Current, the height corresponds to the cur-
rent.

Evaluation parameters for standard

Curve type

Curve type for the evaluation of the calibration curve. Determines the type
according to which the calibration curve is to be fitted. A minimum num-
ber p of different concentrations must exist for the various curve types so
that it is possible to fit this calibration curve. The minimum number p cor-
responds to the number of terms in the calibration function.

Selection

Linear regression | Quadratic regression | Non-
linear regression
| Linear interpolation

Default value

Linear regression

Selected curve type

Calibration function

p

Linear regression

y = CALC 0 + CALC 1

x

≥ 2

Quadratic regres-
sion

y = CALC 0 + CALC 1

x + CALC 2

x

2

≥ 3