3 properties - history – Metrohm viva 1.1 User Manual

Page 831

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

viva 1.1

■■■■■■■■

819

[E-mail...]

[E-mail...] opens the Send e-mail window.

Acoustic signal

on | off (Default value: off)

If this check box is activated, an acoustic signal will be emitted in addition
to the message.

Action
The actions can only be edited if monitoring for common variables is
enabled.

If during common variable monitoring it is found that the validity period
has expired, then one of the following actions will be triggered automati-
cally during the start test:

Selection

Record message | Display message | Cancel
determination

Default value

Display message

Record message
The message that the validity period of the common variable has
expired will be automatically saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued, then the message
that the validity period of the common variable has expired will be
automatically saved with the determination.
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].

6.10.3.3

Properties - History

Tab: Configuration

Common Variables [Edit] Properties... Common varia-

ble - 'Name'

History

History table
The table contains the last 10 values for the selected common variable
and cannot be edited or sorted. The individual values are sorted according
to date so that the most recent assignment is shown last.

Value

The values are shown in the following colors:

Blue, if the value has been assigned automatically by a method.

Black, if the values have been entered manually.

Orange, if the values are outside the warning limits.