3 general - curve evaluation – Metrohm viva 1.1 User Manual

Page 664

Advertising
background image

5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

652

■■■■■■■■

viva 1.1

NOTE

The parameter Reversed peaks is disabled in the CP command,
because there is no peak evaluation for the CP command.

5.6.2.3

General - Curve evaluation

Tab: Method

Evaluation General Curve evaluation

Fixed point evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(or an independent variable) for the fixed point from the measuring point
list. These values can then be accessed via the formula editor and be out-
put as user-defined results.

Table

NOTE

The table has ten lines at most.

The table on the Curve evaluation tab contains an overview of the
parameters defining the fixed point evaluation. The table contains the fol-
lowing columns:

FP

Number of the fixed point for the assignment of the fixed point F

{z} in the

result variables in the formula editor (see Chapter 2.3, page 21).

Measured quantity

Measured quantity (or independent variable) to which the associated val-
ues for the other quantities will be interpolated from the measuring point
list.

Fixed value

Fixed value of the corresponding number of the selected line in the table.

Unit

Unit of the measured quantity (or the independent variable).

Advertising